| Name | TDSscan20 Pocket TDS Tester |
| Code | bante-0085 |
| Type | Standard |
| Brand | Bante |
| Category | TDS Meters |
| subcategory | Pocket Testers |
| Unit | pcs (pcs) |
| In Stock | Yes |
| Berminat hub kami: | |
TDSscan20 Pocket TDS Tester
Multi-range pocket TDS meter, 1 to 3 points calibration, selectable TDS conversion factors, setup menu contains 5 options, measurement range: 0 to 20 ppt.
Features:
• Multi-range pocket TDS meter covers the low, mid, high ranges.
• 1 to 3 points calibration with automatic recognition for TDS standards.
• Auto-read function senses and locks the measurement endpoint.
• Auto-power off effectively conserves battery life.
• Setup menu allows setting the number of calibration points, TDS conversion factor, etc.
• Reset function automatically resumes all settings back to the factory defaults.
• Replaceable electrode module reduces maintenance and replacement cost.
• Waterproof feature to ensure complete protection in harsh environments.
Specification
Model | TDSscan20 | |
TDS | ||
Range | 0 to 10.00, 100.0, 1000 ppm, 20.00 ppt | |
Resolution | 0.01, 0.1, 1, automatic | |
Accuracy | ±1% F.S. | |
Calibration | 1 to 3 points | |
Calibration Solutions | 71.8 ppm / 744.7 ppm / 7447 ppm | |
Temperature | ||
Range | 0 to 60°C / 32 to 140°F | |
Resolution | 0.1°C/°F | |
Accuracy | ±1°C / ±1.8°F | |
Calibration | 1 point | |
Calibration Range | Reading ±10°C/°F | |
Other Specifications | ||
Temperature Compensation | 0 to 60°C / 32 to 140°F, automatic | |
TDS Conversion Factor |
| 0.1 to 1.0 (default 0.5) |
Hold Function | Manual / Auto-read measurement endpoint | |
Auto-Power Off | 8 minutes after last key pressed | |
Display | Dual-line LCD (21 × 21 mm) | |
Power Supply | 2 × 1.5V AAA batteries | |
Battery Life | Approximately 200 hours of continuous use | |
Environmental Conditions | 0 to 50°C / 32 to 122°F, relative humidity < 80% | |
Dimensions | 185 × 40(Ø) mm | |
Weight | 100 g |
Ordering Info
Ordering Code | Tester Includes |
TDSscan20 | TDS standard solutions 71.8 ppm, 744.7 ppm, 7447 ppm |
Carrying case |
Optional Accessories
| Items | Description |
| E-ECSCAN-C1-10K Electrode | Replacement conductivity electrode |
| TDS Standard Solutions | For TDS calibration |
Packaging
Ordering Code | Dimensions and Weights |
TDSscan20 | 28 × 8 × 25 cm, 1.5 kg |