| Name | Microwave Digestion System 12/18 Vessels |
| Code | scitek-0895 |
| Type | Standard |
| Brand | Scitek |
| Category | Microwave Digestion |
| subcategory | Microwave Digestion System 12/18 Vessels |
| Unit | pcs (pcs) |
| In Stock | Yes |
| Berminat hub kami: | |
Microwave Leakage : <5mw/cm2
Temperature Control Accuracy : ±0.3℃
Temperature Controlling Range : 50~400℃
Rotation : 360° Continuous Rotating
Size(W×D×H)(mm) : 640x630x690(Main device)
Microwave Digestion System Specifications:
| Model | MDS-12 | MDS-18 |
|---|---|---|
| Vessel Quantity | 12 Vessels | 18 Vessels |
| Vessel Volume | 100 ml | |
| Temperature Monitoring | Contactless IR sensor | |
| Temperature Control | Scanning control of each vessel | |
| Temperature Control Range | 50~400℃ | |
| Temperature Control Accuracy | ±0.3℃ | |
| Maximum Working Temperature | 250℃ | |
| Inner Vessel Temperature Limit | 300℃ | |
| Display | 7 inches Touch Screen | |
| Rotation Mode | 360° Continuous Rotation | |
| Vessel Material | lnner: lmported TFM | |
| Outer; Imported PEEK+Glass Fiber | ||
| Pressure Testing | Contactless pressure sensor | |
| Maximum Working Pressure | 6 MPa | |
| Pressure Control Range | 0-10MPa | |
| Pressure Control Accuracy | 0.01 MPa | |
| Power Supply | AC 220V, 50/60Hz | |
| Microwave Power | 0-2000W Adjustable/ 0-3000W Adjustable | |
| Microwave Frequency | 2450Hz | |
| Microwave Leak | <5mw/cm2 | |
| Sizes(W×D×H)(mm) | 640x630x690(Main device) | |
| Shipping Size(W×D×H)(mm) | 800x800x810(Main device); 620x590x560(Accessories) | |
| N.W./G.W. (kg) | 70kg(Main device) + 26kg (Accessories) /120kg (Main device) + 31kg (Accessories) | 75kg (Main device) + 33kg (Accessories) /120kg (Main device) + 38kg (Accessories) |
Microwave Digestion System Feature:
Special Designed Sample Digestion Vessel
The automatic vent and self-resealing structure which won national patent ensures the digestion vessels can automatically release pressure and instantly
reseal when a sudden over-pressure situation occurs. It prevents waste of batch samples and solutions and also avoids vessel damage.
Contact-less Temperature and Pressure Monitoring System
The advanced contact-less IR sensor could measure the real-time temperature of sample solutions in each digestion vessel. Meanwhile, the contact-less pressure sensor could monitor the real-time pressure of each vessel. This realizes isolation between temperature and pressure controlling system, also avoids sample cross-contamination. Each real-time temperature and pressure value is displayed during the whole digestion process, allowing a clear check of digestion conditions.
Large Storage Capacity
It can edit and store 255 types of programs according to user's requirements and each method program can set parameters (temperature, pressure, time,microwave power) according to their own requirements which provides convenient operation for users.
Safety Protection System
With high-strength double locked security door, the real-time temperature and pressure monitoring, automatic adjustment of overpressure and over-temperature and abnormal sound monitoring, the device can run in a highly safe environment.
Power Adjustment
The variable frequency resonance or non-pulse(optional) continues rotating mode ensures high efficient and even microwave throughput. It can be set
between 0 ~2000W/0~3000W according to user's requirements.
Microwave Digestion System Detail Show

Microwave Digestion Vessel Structure

MDS-12 Inner Structure

MDS-18 Inner Structure
Arabic
French
German
Indonesian
Italian
Portuguese-brazilian
Simplified-chinese
Spanish
Thai
Traditional-chinese
Turkish
Vietnamese