Double Beam UV VIS Spectrophotometer SP-DUV9000 Series (scitek-0421) Products
Name Double Beam UV VIS Spectrophotometer SP-DUV9000 Series
Code scitek-0421
Type Standard
Brand Scitek
Category Spectrophotometer and Colorimeter
subcategory Double Beam UV-Vis Spectrophotometer
Unit pcs (pcs)
In Stock Yes
Berminat hub kami: wa chat
Product details for Invoice
Model : SP-DUV9000; SP-DUV9000I; SP-DUV9000II

Wavelength Range : 190-1100nm
Wavelength Accuracy : ±0.1nm(D2 656.1nm), ±0.3nm@all
Wavelength Repeatability : 0.2nm
Size(W×D×H)(mm) : 625x430x206
Product Details

UV VIS Spectrophotometer Features:

SP-DUV9000 series are wide-screen dual-beam spectrophotometers. They are the best choice for high quality spectrophotometers.

  1. Double beam design prevents circuit fluctuation and stray light and ensures the instrument's reliability.

  2. Can operatee many functions directly on PC.

  3. A 5200mm long optical path greatly improves the resolution.

  4. 16mm optical base ensures the stability of the instrument.

  5. All baselines, wavelengths and dark current can be automatically calibrated.

  6. Equipped with a USB port to connect with the PC, easy to realize data transmission.

UV VIS Spectrophotometer Specifications:

ModelSP-DUV9000SP-DUV9000ISP-DUV9000II
Optical SystemDouble Beam, 1200 Lines/mm Grating)
Wavelength Range190-1100nm
Bandwidth1.8nm1.0nm0.5/ 1.0/ 2.0/ 4.0 nm
Wavelength Accuracy±0.1nm(D2 656.1nm), ±0.3nm@all
Wavelength Repeatability0.2nm
Photometric Accuracy±0.2%T(0-100%T)
Photometric Repeatability≤0.1%T(0-100%T)
Photometric Range-0.3-3A
0-200%T
0-9999C
Stability± 0.001A/h @ 500nm
Baseline Flatness± 0.001A/h
Noise± 0.0005A/h
Stray Light≤0.03%T @ 220nm,360nm
Data Output PortUSB
Printer PortParallel Port
Display320x240 Dots LCD
LampsTungsten Lamp & Deuterium Lamp
DetectorSilicon Photodiode
Power RequirementsAC 220V/50Hz or 110V/60Hz
Size(W×D×H)(mm)625x430x206
Shipping Size (W×D×H)(mm)850x650x450
N.W./G.W. (kg)30/40