Total Organic Carbon Analyzer, NDIR Detector (scitek-1130) Products
Name Total Organic Carbon Analyzer, NDIR Detector
Code scitek-1130
Type Standard
Brand Scitek
Category Environment Monitor
subcategory TOC Analyzer
Unit pcs (pcs)
In Stock Yes
Berminat hub kami: wa chat
Product details for Invoice
Model : TOC-A200
Measurement Range : 0-30000mg/l (can exanded)
Repeatability : 3%
Maximum Salinity : 85g/l
Detection Limit : 50µg/l
Gas Requirement : Oxygen, Nitrogen ≥ 99.995%
Size(W×D×H)(mm) : 455x430x440
 
Product Details

Operation Principle:

The UV lamp induces oxidation in the organisms, transforming organic substances into carbon dioxide. The detection method employed is the direct conductivity method. Total organic carbon (TOC) is calculated as the disparity between the total carbon (TC) concentration in samples tested after oxidation and the concentration in samples not subjected to oxidation, which accounts for the total inorganic carbon (TIC). In other words, TOC is calculated as TOC = TC - TIC.

Total Organic Carbon Analyzer Features:

  1. Our proprietary signal management system, protected by patent law, offers significant advantages such as online configuration, real-time monitoring, self-diagnostic capabilities, and flow speed control, ensuring optimal performance and safe operation.

  2. The low-current system design is also geared towards enhancing operational safety.

  3. Varied temperature settings for different samples guarantee thorough sample digestion, enhancing the accuracy of measurement data.

  4. Dynamic adjustment of cooling module power based on sampling volume enhances drying performance, ensuring the delivery of dry gas to the detector.

  5. The automatic leak check system mitigates operational errors, boosting instrument performance and overall safety during operation.

  6. The flow rate control system prevents any adverse effects caused by fluctuations, ensuring precise data.

  7. TOC detector, featuring a 24-bit data solution, expands the monitoring range, while the controlling system, utilizing 32-bit processing technology, significantly enhances overall performance.

Application of A200 TOC Analyzer:

Widely used in drinking water, industrial water, sewage, wastewater, and monitoring of rivers, lakes and oceans, surface water etc.

Total Organic Carbon Analyzer Specifications:

ModelTOC-A200
Measurement Range0-30000mg/l (Can exanded)
Repeatability3%
Maximum Salinity85g/l
Detection Limit50µg/l
Gas RequirementOxygen, Nitrogen ≥ 99.995%
DetectorNDIR
ParametersTC, TIC, TOC, NPOC
Digestion ModeHigh Temperature Combustion
ApplicationLiquid Sample
Audit Trail FunctionOptional
Power SupplyAC100-240V, 50/60Hz
Size(W×D×H)(mm)455x430x440
Shipping Size(W×D×H)(mm)850x840x730
G.W. (kg)65