| Name | Horizontal Pressure Steam Autoclave 150/200/300/400L |
| Code | scitek-0117 |
| Type | Standard |
| Brand | Scitek |
| Category | Autoclaves & Sterilizers |
| subcategory | Horizontal Autoclave |
| Unit | pcs (pcs) |
| In Stock | Yes |
| Berminat hub kami: | |
Capacity (L) : 150 / 200 / 300 / 400
Chamber Size (ØxH)(mm) : Ø472x880 / Ø472x1150 / Ø620x995 / Ø620x1320
Working Temperature (℃) : 105-136
Working Pressure (Mpa) : 0.225
The horizontal design of this pressure steam autoclave makes loading and removing items easier and reduces operational labour.
Horizontal Pressure Steam Autoclave Specifications:
| Model | ST-H150I | ST-H200I | ST-H300I | ST-H400I |
|---|---|---|---|---|
| Volume (L) | 150 | 200 | 300 | 400 |
| Steam Consumption (KG) | ≤12 | ≤13 | ≤18 | ≤22 |
| Water Consumption (KG) | ≤30 | ≤30 | ≤40 | ≤40 |
| Power (KW) | ≤2+⑮ | ≤2+⑮ | ≤2+⑱ | ≤2+⑱ |
| Chamber Size (ØxL)(mm) | Ø472x880 | Ø472x1150 | Ø620x995 | Ø620x1320 |
| Sizes (WxDxH)(mm) | 1290x690x1680 | 1560x690x1680 | 1430x850x1760 | 1755x850x1760 |
| Shipping Size (WxDxH)(mm) | 1570x840x1850 | 1740x840x1850 | 1610x990x1900 | 1935x990x1900 |
| N.W./G.W. (kg) | 400/500 | 450/550 | 510/610 | 650/750 |
| Design Pressure | -0.1~0.25Mpa | |||
| Design Temperature | 139℃ | |||
| Working Pressure | 0.225Mpa | |||
| Working Temperature | 105-136℃ | |||
| Temp Accuracy | ±1℃ | |||
| Temperature Equilibrium | <+2.0℃ | |||
| Water Pressure | 0.1-0.2Mpa | |||
| Compressed Air Pressure | 0.4~0.7Mpa | |||
| Steam Pressure | 0.3-0.7Mpa | |||
| Power Voltage | Three-phase five-wire AC 380V 50Hz | |||
Horizontal Pressure Steam Autoclave Features:
Horizontal circular inner chamber and jacket structure for safety, heat preservation and drying function.
Control system featuring HMI man-machine interface, touch screen and microcomputer for intuitive operation.
Disc type embedded gear and electric lock door structure for reliable and secure operation.
Self-expanding door seals for effective sealing.
The horizontal pressure steam autoclave equipped with a standard equipment verification interface for easy calibration and validation.
Optimal design pipeline system with SUS 304 pipeline and clamp structure for quick assembly.
FO value and temperature-time double guarantee sterilization effect with comprehensive sterilization records.
Multiple safety protections including over-temperature and over-pressure protection.
Built-in printer & USB function.
Arabic
French
German
Indonesian
Italian
Portuguese-brazilian
Simplified-chinese
Spanish
Thai
Traditional-chinese
Turkish
Vietnamese