Speed Range(rpm) : 200-15000
Max RCF (xg) : 21380
Temperature Setting (℃) : -20~+40
Timer : 30s-99 min
Capacity Rnage : 1.5/2mL×24, 0.5mL×36, PCR8×4, 5mL×12, 5mL×18
Features a simple, intuitive digital interface and durable qualities for years of use.
The CFG-T15HR laboratory refrigerated centrifuge is uniquely designed to meet various applications in clinical and research laboratories. The centrifuge is available with optional adapters for all common blood collection and culture tube sizes. Unlike most compact centrifuges, the rotor can be easily removed from the centrifuge chamber for cleaning.
Benchtop Refrigerated Centrifuge Features:
Ergonomic electronic lock design for easy and quick closing of the centrifuge lid.
Pre-cooling feature that automatically starts the compressor to cool the chamber when the set temperature is lower than the ambient temperature.
Sound indicator to alert when the upper cover is closed to maintain the temperature inside the chamber after centrifugation.
Use of advanced R290 fluorine-free refrigerant for low carbon and environmental protection.
Temperature range of -20℃ to 40℃ with fast and powerful cooling capacity, allowing for cooling from room temperature to 4℃ in just 7 minutes, and maintaining temperature at maximum speed.
Benchtop Refrigerated Centrifuge Specifications:
| Model | CFG-T15HR |
|---|---|
| Max. Speed | 15,000rpm (200-15,000rpm) in 100rpm increments. |
| Max. RCF | 21,380×g, step: 10×g |
| Capacity | 1.5/2mL×24, 0.5mL×36, PCR8×4, 5mL×12, 5mL×18. |
| Temperature Setting | -20℃- 40℃ |
| Run Time | 30s-99 min; HOLD (Continuous running) |
| Acceleration and Deceleration Time | 25s↑25s↓ |
| Motor Type | DC brushless motor |
| Safety Performance | Dual door lock, over-speed detection, over-temperature detection with Internal diagnosis system. |
| Power Requirements | Single phase, 200-240V, 50Hz, 500W |
| Noise | ≤56dB |
| Sizes(W×D×H)(mm) | 553x332x283 |
| Shipping Size(W×D×H)(mm) | 900x500x600 |
| N.W./G.W. (kg) | 30/39 |
| Other Functions | Rotation speed/RCF switchover, inching operation, running process indicator, sound alert. Acceleration/deceleration in 9 steps, program storage capacity. |
Advantages of benchtop refrigerated centrifuge:
Storage capacity for up to 9 programs to meet experimental needs.
Three-layer sealed rotors to ensure air-tightness and biological safety.
High-strength aluminum alloy construction capable of withstanding numerous high temperature and high-pressure sterilizations.
Five rotors with varying capacities for centrifuge tubes and PCR tubes (0.2mL, 0.5mL, 1.5mL, 2mL, and 5mL).
External condensation tank to prevent water accumulation within the centrifuge chamber.
Rotor imbalance detection with an alarm for added safety.
Applications of benchtop refrigerated centrifuge:
Molecular Biology: Isolating DNA, RNA, and proteins from cell lysates.
Biochemistry: Purifying and concentrating enzymes, antibodies, and other biomolecules.
Clinical Diagnostics: Processing blood, urine, and other biological samples for analysis.
Cell Biology: Separating cells and subcellular organelles such as mitochondria and nuclei.
Pharmaceuticals: Preparing samples for drug formulation and stability testing.