Automatic Elisa Plate Reader, 96 Wells Plate (scitek-0825) Products
Name Automatic Elisa Plate Reader, 96 Wells Plate
Code scitek-0825
Type Standard
Brand Scitek
Category Elisa Reader and Washer
subcategory Elisa Microplate Reader
Unit pcs (pcs)
In Stock Yes
Berminat hub kami: wa chat
Product details for Invoice
Model : ELMR-96; ELMR-96I
 
Wavelength Range : 340~750nm
Repeatability : [0, 3.0)≤0.3%, [3, 4.0)≤1%
Resolution : 0.001Abs
Accuracy : [0, 2.0)≤ ± 0.005, [2.0 ,3.0)≤ ± 1%, [3.0, 4.0)≤ ± 1.5%
Sizes(W×D×H)(mm) : 280x432x220
The elisa microplate reader is a fast and accurate way to read absorbance data from microplates and is suitable for use in biomedical research, clinical diagnostics, drug discovery and more.
Product Details

Elisa Plate Reader Specifications:

Model ELMR-96/ELMR-96I
Light Source 6V 10W halogen lamp
Wavelength Range 340~750nm
Filter Standard configuration: 405, 450, 492, 630nm four filters; It can be loaded with up to eight filters
Absorbance Range 0.000-4.000Abs
Resolution 0.001Abs
Linearity (0, 2.0)≤±1%; [2.0, 4.0)≤±2%
Wavelength Indication Error ≤±3nm
Wavelength Repeatability ≤±1.5nm
Repeatability [0, 3.0)≤0.3%, [3, 4.0)≤1%
Stability [0, 3.0)≤0.3%, [3.0, 4.0)≤1%
Accuracy [0, 2.0)≤ ± 0.005, [2.0 ,3.0)≤ ± 1%, [3.0, 4.0)≤ ± 1.5%
Sensitivity Sensitivity≥0.010
Channel Difference <0.01A(1.0Abs standard filter )
Measuring Speed Single wavelength <10s/96 wells, dual wavelengths <16s/96 wells (normal mode)
Incubation Temperature Room temperature+4℃-50℃ (ELMR-96I)
Ambient Temperature 4℃~ 45℃
Relative Humidity ≤70%
Power Supply AC100-240V, 50/60Hz
Sizes(W×D×H)(mm) 280x432x220
Shipping Size(W×D×H)(mm) 580x430x360
N.W./G.W. (kg) 8.5/13

Elisa Plate Reader Features:

The ELMR-96/ELMR-96I is a popular elisa plate reader capable of sensitively measuring absorbance signals on microtiter plates, accurately detecting and analyzing the concentration and activity of biomolecules.

  1. 7-inch high-resolution LCD capacitive touch screen for simple operation and easy viewing of test results

  2. Detection accuracy with single wavelength and dual wavelength detection for fast plate reading

  3. A wavelength range of 340-750nm can meet a variety of applications

  4. 8-bit filter wheel, 4 filters as standard, with various other optional filters

  5. With vibration plate function, vibration plate time and speed adjustable

  6. With self-test and diagnostic functions for optical path, moving position, etc

  7. Light source energy-saving design maximises the service life of the light source

  8. There is no need to open the shell design; you can complete the replacement of light sources and filters

Details of 96 well microplate reader:

elisa plate reader

elisa plate reader

Power connector and data transfer port for easy connection of the 96 well microplate reader to power supply and external computer.

elisa microplate reader

This elisa reader is equipped with a high-definition color display for intuitive viewing of test results and device status.

ELMR-96I (2)

ELMR-96I (5)

Absorbance assay:

Working Principle: Absorbance detection is used to quantitatively analyze the amount of biomolecules in a sample by measuring the light absorption value of the sample at a specific wavelength. When light passes through the sample, the sample absorbs a certain amount of light energy, and the remaining light energy is measured by the detector to calculate the absorbance value of the sample.

Application Scenario: Absorbance assay is widely used in enzyme-linked immunosorbent assay (ELISA) technology to determine the concentration of biological molecules such as nucleic acids and proteins. In addition, it is used in biological studies such as cell proliferation and toxicity detection.